5 0
USER INTERFACE
ELECTROTHERAPY OPERATION
Complete the following steps to begin Electrotherapy treatment:
1. Prepare patient and therapy system for Electrotherapy. Refer to the PATIENT PREPARATION section on for
electrode selection, preparing the patient, and securing electrodes.
2. Select ELECTROTHERAPY icon from the home screen
3. Select desired waveform
Note: Refer to the System Specifications section of this manual for all waveform specifications for the Intelect
4. SET UP TREATMENT
On the treatment review screen - you can adjust treatment parameters to desired level.
Never start with intensity adjustment – first adjust all other parameters and set Intensity just before starting treatment
Pressing the save button will save the treatment as a custom protocol which can be named by the user
Pressing the Assign to button will give two buttons
-Assign to: this button assigns the treatment data to a specific treatment data file which can be named by the
user
-Open pain scale: this button opens up the pain scale so the pre treatment pain can be recorded
5. START TREATMENT
Press the start/pause button
7. PAUSE TREATMENT
Press the Start/Pause button
INTELECT
MOBILE 2 USER MANUAL
®
E N
Touch to select
Swipe horizontally to
see more
Mobile 2.
®