32
SYSTEM
WAVEFORMS (CONTINUED)
SURGED: Monophasic Rectangular
The SURGED: Monophasic Rectangular Current is a
series of rectangular, monophasic pulses. The pulses
surge to maximum power, hold and then decrease
before the pause. This waveform is well suited for
muscle re-education.
Output Mode
Electrodes
Available on Channels
1, 2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0- 60 mA (CC)
Frequency
5-60 Hz
Phase Duration
0.2-5.0 ms
Pause
0-57 sec
Surges per minute
1-20
IRMS
0-37mA
DC component
Yes
Low Level Galvanic
Advice on size and type of electrodes to be used is
given in device GUI "treatment guidelines" feature
Low Level Galvanic Current is a direct current
flowing in one direction only. The intensity is limited
to 4.0mA
Output mode
Electrodes
Mode selection
CC
Output Intensity
0-4 mA (CC)
Dosage
40-80 mA-min
Polarity
Fixed at positive
IRMS
0-5mA
DC component
Yes
INTELECT
MOBILE 2 USER MANUAL
®
SURGED:Monophasic Triangular
Surged Monophasic Triangular Pulsed waveform is
a one channel waveform. It is a triangular pulse
waveform that is ramped up and down in amplitude
(surged).
Output mode
Electrodes
Available on channels
1,2
treatment time
1-60min
Mode selection
CC
Output intensity
0-60mA(CC)
Frequency
5-60Hz
Phase duration
0.2-5.0ms
Pause
0-57sec
surges per minute
1-20
IRMS
21mA
DC component
Yes
VMS™ FR
The VMS-FR version of the VMS waveform is a
physiologically based channel interaction in which
one channel stimulates the agonist and the other
the antagonist of the muscle group that is being
exercised. The agonistic channel initiates the
movement with a brief burst of power, followed
by a period of sustained activity to complete the
movement. The antagonistic channel has a brief
burst of power to slow down the initial acceleration
of the agonist, followed by a low output to regulate
the movement of the agonist. The movement is
completed by a final burst of activity in both
channels. VMS is a symmetrical biphasic waveform
with a 100 µsec interphase interval. Because the
pulse is relatively short, the waveform has a low skin
load, making it suitable for applications requiring
high intensities, such as in muscle re-education
protocols.
Output Mode
Electrodes
Available on
Channels 1&2
Treatment Time
1-60 min
Mode Selection
CC
Output Intensity
0-140 mA (CC)
Burst Duration
200 - 5,000 ms
Cycle Time
TBD
Frequency
20-80 pps
Phase Duration
30-400 µsec
IRMS
0-39mA
DC component
No
E N