Live Mode
Live mode - allows the enabled output to follow the state of the inhibit signal. When the inhibit signal is
true, the output is disabled; when the inhibit signal is false, the output is enabled. The inhibit signal is
not latched, and the inhibit event is not treated as a protection event. Therefore, a protection clear is
not required when switching between inhibit true and inhibit false.
Step 1. Turn the AC source off
Step 2. Connect an external switch or circuit to the Inhibit terminals.
Step 3. Turn the AC source on.
Step 4. Select Live as the inhibit mode.
SCPI:
OUTPut:INHibit:MODE
Step 5. Select a polarity for the Live inhibit mode.
SCPI:
DIGital:PIN:POLarity
Step 6. Turn the output on. When the inhibit signal disables the output, the status indicator displays
"INH". The voltage and current meters will display zero.