Télécharger Imprimer la page

Keysight AC6800B Serie Guide D'utilisation Et D'entretien page 114

Publicité

Les langues disponibles

Les langues disponibles

3  SCPI Programming Reference
IEEE-488 Subsystem
Command/Query Description
*CLS
Clear status command.
*ESE <value>
Event status enable command and query.
*ESE?
*ESR?
Event status event query.
*IDN?
Identification Query. Returns the instrument's identification string.
*LRN?
Learn query. Returns the SCPI strings that reproduce the present instrument settings.
*OPC
Sets the OPC (operation complete) bit in the standard event register.
*OPC?
Returns a 1 to the output buffer when all pending operations complete.
*OPT?
Returns a string identifying any installed options.
*PSC 0|1
Enables (1) or disables (0) the clearing of certain enable registers at power on.
*PSC?
*RCL <0 to 10>
Recalls a saved instrument state.
*RST
Resets the instrument to default values.
*SAV <0 to 10>
Saves the instrument state to a nonvolatile memory location.
*SRE <value>
Service request enable command and query.
*SRE?
*STB?
Status byte query. Reads the Status Byte Register, which contains the status summary bits and the Output
Queue MAV bit.
*TRG
Trigger command. Applies a software trigger (equivalent to IEEE 488.1 Device Trigger) for both TRANsient
and ACQuire trigger groups.
*TST?
Self-test. Returns the errors found during the most recent power-on self-test.
INITiate Subsystem
Command/Query
INITiate[:IMMediate]:ACQuire
INITiate[:IMMediate]:TRANsient
INITiate:CONTinuous:ACQuire ON|1|OFF|0 Initiates a new measurement and enables or disables "continuous" mode.
LXI Subsystem
Command/Query
LXI:IDENtify[:STATe] ON|1|OFF|0
LXI:IDENtify[:STATe]?
114
Description
Initiates the ACQuire (measurement) operation.
Initiates the TRANsient operation.
Description
Turns the front-panel LXI identify indicator (blinking "Lan" annunciator) on or off.