Télécharger Imprimer la page

Keysight AC6800B Serie Guide D'utilisation Et D'entretien page 77

Publicité

Les langues disponibles

Les langues disponibles

VOLTage:LIMit:UPPer MAX
VOLTage:LIMit:LOWer MIN
VOLTage 120
FREQuency:LIMit:UPPer MAX
FREQuency:LIMit:LOWer MIN
FREQuency 70
The commands above set the limits to accommodate the widest possible VOLTage and FREQuency
settings within the given range. On the 155-V voltage range, the VOLTage setting cannot exceed 157.5
V.
In AC+DC mode, the voltage can be specified only when the AC and DC voltage settings are within the
voltage limit range and the peak of the AC+DC waveform is between -445 and 445 V.
Keysight AC6800B Series Operating and Service Guide
Sets upper limit to maximum, 315 V
Sets lower limit to minimum, 0 V
Sets voltage to 120 V
Sets upper limit to maximum, 500 Hz
Sets lower limit to minimum, 40 Hz
Sets frequency to 70 Hz
2  User Information
77

Publicité

loading

Ce manuel est également adapté pour:

Ac6801bAc6802bAc6803bAc6804b