CURRent Subsytem
The CURRent subsystem limits the instrument's output current.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <value>|MINimum|MAXimum
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MINimum|MAXimum]
Sets the immediate AC current limit in amps (rms).
Parameter
Typical Return
AC6801B 0.1 to 5.2 A
+9.00000E-01
AC6802B 0.2 to 10.5 A
AC6803B 0.4 to 21.0 A
AC6804B 0.8 to 42.0 A
Sets the current limit to 900 milliamps: CURR 900 MA
Current units (UA, MA, A) are allowed.
Values exceeding the instrument's maximum value will be set to the instrument's maximum value.
The
*RST
default is the instrument's MAXimum value.
[SOURce:]CURRent:OFFSet[:IMMediate] <value>|MINimum|MAXimum
[SOURce:]CURRent:OFFSet[:IMMediate]? [MINimum|MAXimum]
Sets the immediate DC current limit in amps.
Parameter
Typical Return
AC6801B 0.1 to 4.2 A
+3.00000E+00
AC6802B 0.2 to 8.4 A
AC6803B 0.4 to 16.8 A
AC6804B 0.8 to 33.6 A
Set the DC current limit to 3 A: CURR:OFFS 3
Current units (UA, MA, A) are allowed.
Values exceeding the instrument's maximum value will be set to the instrument's maximum value.
The
*RST
default is the instrument's MAXimum value.
Keysight AC6800B Series Operating and Service Guide
3 SCPI Programming Reference
125