Télécharger Imprimer la page

Keysight AC6800B Serie Guide D'utilisation Et D'entretien page 170

Publicité

Les langues disponibles

Les langues disponibles

3  SCPI Programming Reference
SCPI Error Messages
The error messages below are in the following format, with a numeric error code and a quoted error
string separated by a comma:
     -222,"Data out of range"
The error/event queue is empty and returns 0,"No error" when the *CLS common command is sent, the
last item in the error queue is read, or the instrument cycles power.
Error Code
Category Meaning
Range
-100 to -199
Command
errors
-200 to -299
Execution
errors
-300 to -399
Device-spe-
cific errors
-400 to -499
Query
errors
100 to 399
Specific
SCPI Errors
900 and above Self-test
errors
Error Code Error Text
-101
Invalid character
-102
Syntax error
-103
Invalid separator
-108
Parameter not allowed
-109
Missing parameter
-110
Command header error
-112
Program mnemonic too long
-113
Undefined header
-114
Header suffix out of range
-115
Unexpected number of parameters
-120
Numeric data error
-128
Numeric data not allowed
-131
Invalid suffix
-138
Suffix not allowed
170
An IEEE 488 syntax error has been detected by the instrument's parser.
An error has been detected by the instrument's execution control block.
A device-specific error has been detected, possibly related to a hardware failure.
The instrument's output queue control has detected a problem with the message
exchange protocol.
An error in the instrument's operation has occurred, such as a settings conflict or
an attempt to execute a command when the instrument is in an incorrect mode.
A self-test error has occurred.
Keysight AC6800B Series Operating and Service Guide
Bit in Event
Status
Register
(n/a)
(n/a)
(n/a)
(n/a)
(n/a)
2

Publicité

loading

Ce manuel est également adapté pour:

Ac6801bAc6802bAc6803bAc6804b