3  SCPI Programming Reference
STATus Subsystem
Status register programming lets you determine the operating condition of the instrument at any time.
The instrument has three groups of status registers; Operation, Questionable, and Standard Event. The
Operation and Questionable status groups each consist of the Condition, Enable, and Event registers
as well as NTR and PTR filters.
The Status subsystem is also programmed using Common commands. Common commands control
additional status functions such as the Service Request Enable and the Status Byte registers. Refer to
Status Overview
for more information.
STATus:OPERation[:EVENt]
Queries the
event register
(latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation
Status Event register clears it.
Parameter
Typical Return
(none)
<bit value>
Read the operation status event register: STAT:OPER?
The value returned is the binary-weighted sum of all bits set in the register.For example, with bit 3
(value 8) and bit 5 (value 32) set (and corresponding bits enabled), the query returns +40.
*RST
does not affect this register unless the reset causes an event that the filters are configured to
capture.
STATus:OPERation:CONDition?
Queries the operation condition register. This is a read-only register, which holds the instrument's live
(unlatched) operational status. Reading the Operation Status Condition register does not clear it.
Parameter
(none)
Read the operation status condition register: STAT:OPER:COND?
The value returned is the binary-weighted sum of all bits set in the register.For example, with bit 3
(value 8) and bit 5 (value 32) set (and corresponding bits enabled), the query returns +40.
The condition register bits reflect the current condition. If a condition goes away, the corresponding bit
is cleared.
A
*RST
clears this register, other than those bits where the condition still exists after *RST.
150
for the
Operation Status
Typical Return
<bit value>
Keysight AC6800B Series Operating and Service Guide
group. This is a read-only register, which stores