4  Calibration, Verification, and Service
End Calibration
This section describes the final steps in the calibration process.
Storing calibration constants overwrites the existing ones in nonvolatile memory. If you do
not save the calibration data, the new calibration constants will only be used until you exit
the calibration state or cycle power. For the front panel, leaving the calibration state is
defined as exiting the menu levels at or below [Menu] > System > Admin > Cal. For SCPI,
leaving the calibration state is accomplished by CAL:STAT OFF.
If you are not sure you want to permanently store the new constants, do not save the data
when you exit the calibration mode.
Step
Description
1
Enter the calibration
date.
2
Save calibration data.
3
Log out.
180
Front Panel
[Menu] > System > Admin > Cal > Date
Enter the calibration date in the Data field. Press
[Select] or [Enter] to finish.
[Menu] > System > Admin > Cal > Save
[Menu] > System > Admin > Logout
Keysight AC6800B Series Operating and Service Guide
SCPI
CAL:DATE "<
string>"
CAL:SAVE
CAL:STAT OFF