3  SCPI Programming Reference
Standard Event Status Group
These registers are programmed by Common commands. The group consists of an Event and Enable
register. The Standard Event event register latches events relating to communication status. It is a
read-only register that is cleared when read. The Standard Event enable register functions similarly to
the enable registers of the Operation and Questionable status groups. See
description of each register.
The event status register bits are set when certain events occur during instrument operation.
The register is controlled by the IEEE 488 commands *ESE, *ESE?, and *ESR?.
Bit Value Bit Name
0
1
Operation Com-
plete (OPC)
1
2
(not used)
2
4
Query Error (QYE) The instrument tried to read an empty output buffer, a command was received before a previous
3
8
Device Dependent
Error (DDE)
4
16
Execution Error
(EXE)
5
32
Command Error
(CME)
6
64
Reserved
7
128 Power ON (PON)
108
Description
All commands before and including *OPC have completed.
0 is returned
query was read, or the input and output buffers are full.
A device-specific error occurred.
An execution error occurred.
Error
rectly depending on the instrument's conditions.
A command syntax error occurred.
0 is returned
Power has been cycled since the last time the event register was read or cleared.
Keysight AC6800B Series Operating and Service Guide
Status Registers
Error Messages
Messages. A valid SCPI command may not be executed cor-
Error Messages
for a