4 Calibration, Verification, and Service
Step Action
4
Program output to RANGE 310, VOLT 0, OUTP:COUP DC, OFFSET 128.
5
Record DVM reading and calculate current.
188
Normal Result
Output voltage at 128 VDC, CV annunciator
on, output current near:
2 A for AC6801B
4 A for AC6802B
8 A for AC6803B
16 A for AC6804B
Readings within specified dc current read-
back limits.
Keysight AC6800B Series Operating and Service Guide