4 Calibration, Verification, and Service
RMS Current Measurement Accuracy
This test verifies the measurement accuracy of the rms current readback.
Step Action
1
Turn off the UUT and connect the appropriate load resistor, current
shunt and DMM as shown above.
2
Turn on the UUT. Program the output to RANGE 155 V,VOLT 80, FREQ
60.
3
Record DMM reading and calculate rms current
4
Program output to RANGE 155 V,VOLT 80, FREQ 500.
5
Record DMM reading and calculate rms current.
6
Turn on the UUT. Program the output to RANGE 310 V,VOLT 160, FREQ
60.
7
Record DMM reading and calculate rms current
8
Turn on the UUT. Program the output to RANGE 310 V,VOLT 160, FREQ
500.
9
Record DMM reading and calculate rms current.
186
Normal Result
Output Voltage at 80 Vrms, CV annunciator on,
output current near:
5 A for AC6801B
10 A for AC6802B
20 A for AC6803B
40 A for AC6804B
Readings are within specified current limits.
Output Voltage at 80 Vrms, CV annunciator on,
output current near:
5 A for AC6801B
10 A for AC6802B
20 A for AC6803B
40 A for AC6804B
Readings are within specified current limits.
Output Voltage at 160 Vrms, CV annunciator on,
output current near:
2.5 A for AC6801B
5 A for AC6802B
10 A for AC6803B
20 A for AC6804B
Readings are within specified current limits.
Output Voltage at 160 Vrms, CV annunciator on,
output current near:
2.5 A for AC6801B
5 A for AC6802B
10 A for AC6803B
20 A for AC6804B
Readings are within specified current limits.
Keysight AC6800B Series Operating and Service Guide