2 User Information
Specifying Output Coupling
The instruments support three types of output coupling:
Coupling Description
AC+DC
AC output with positive or negative DC offset
AC
AC output
DC
Positive or negative (bipolar) DC output
In AC+DC mode the combined peak voltage must be between -222.5 and +222.5 V for the low range or
-445 and +445 V for the high range.
Turn the output off to change the output coupling. The instrument performs error checking to ensure
that pending active settings remain within range.
Front Panel: [Menu] > Output > Coupling
SCPI:
OUTPut:COUPling AC|DC|ACDC
72