3  SCPI Programming Reference
TRIGger Subsystem
The TRIGger subsystem sets the measurement operation and trigger functions for various instrument
operations:
TRIGger:ACQuire
Triggers measurement acquisition.
TRIGger:SYNChronize Synchronizes output phase using OUTPut ON instead of a software trigger.
TRIGger[:TRANsient]
Triggers synchronizes output changes.
TRIGger:ACQuire[:IMMediate]
Sends a software trigger to the ACQuire subsystem.
Parameter
Typical Return
(none)
(none)
Configure and trigger the ACQuire subsystem:
TRIG:ACQ:SOUR BUS
INIT:ACQ
TRIG:ACQ
TRIGger:ACQuire:SOURce IMMediate|BUS
TRIGger:ACQuire:SOURce?
Sets the trigger source that starts the measurement after INIT:ACQ. Specify IMMediate to start the
measurement immediately, or BUS (the default) to wait for a software trigger (*TRG, TRIG:ACQ, or
TRIG:ACQ).
Parameter
Typical Return
IMMediate|BUS IMM or BUS
Set the trigger source to BUS:
TRIG:ACQ:SOUR BUS
This parameter is set to BUS at power-on or after *RCL or *RST.
160