Programming AC Output
You can program the voltage, frequency, frequency limits, and starting phase angle of the AC output.
Before programming the output, select the appropriate output mode and voltage range.
Output Mode
Front Panel: [Menu] > Output > Voltage > AC
SCPI:
OUTPut:COUPling AC
Voltage Range
Front Panel: [Menu] > Output > Range
SCPI:
[SOURce:]VOLTage:RANGe[:UPPer] 155|310|MINimum|MAXimum
The output must be off to change the range settings. The instrument checks to ensure that currently
active settings remain within the new range.