TRIGger:SYNChronize:SOURce IMMediate|PHASe
TRIGger:SYNChronize:SOURce?
Sets the output-on phase control when OUTPut ON is sent.
Parameter
Typical Return
IMMediate|PHASe IMM or PHAS
Enable output-on phase control:
TRIG:SYNC:SOUR PHASE
This parameter is set to IMMediate at power-on or after *RCL or *RST.
TRIGger:SYNChronize:PHASe[:ON] <value>|MINimum|MAXimum
TRIGger:SYNChronize:PHASe? [MINimum|MAXimum]
Sets the phase angle of the output-on phase control in degrees.
Parameter
An integer from 0 to 359 degrees +1.80000E+02
Set the OUTPUT on phase angle to 180°:
TRIG:SYNC:PHAS 180
This command supports the parameter suffix DEG (degrees), but not RAD (radians).
The
*RST
command sets this parameter to 0.
TRIGger:TRANsient[:IMMediate]
Triggers the TRANsient subsystem.
Parameter
Typical Return
(none)
(none)
Configure and trigger the TRANsient subsystem:
VOLT:MODE STEP
FREQ:MODE STEP
TRIG:TRAN:SOUR BUS
INIT:TRAN
TRIG:TRAN
Keysight AC6800B Series Operating and Service Guide
Typical Return
3 SCPI Programming Reference
161