1 Getting Started
Remote Interface Configuration
USB Configuration
GPIB Configuration
LAN Configuration
Modifying the LAN Settings
Using the Web Interface
Using Telnet
Using Sockets
Using HiSLIP
This section describes how to configure each remote interface.
This instrument supports remote interface communication over GPIB (optional), USB, and LAN
(default). All three interfaces are "live" at power up, and they may be used simultaneously. To use these
interfaces, install the Keysight IO Libraries software from the Keysight Automation-Ready CD and
connect the instrument to your PC.
The front-panel IO annunciator indicates remote interface activity. The Lan annunciator appears when
the LAN port is connected and configured. This instrument continually monitors the instrument's LAN
port and automatically reconfigures it when the instrument is disconnected and then reconnected to a
network.
The instrument ships with an Automation-Ready CD that contains Keysight IO Libraries Suite software,
which must be installed to enable remote-interface operations. The CD automatically starts and
provides information on installing the software. The CD also includes the Keysight Technologies
USB/LAN/GPIB Connectivity Guide, which contains additional information.
To begin configuring the remote interface from the front panel, log in to the Admin menu by pressing
[Menu] > System > Admin > Login. The default password is blank. Then press
[Menu] > System > Admin > IO to enable the desired interfaces.
Then press [Menu] > System > IO to configure the individual interfaces.
The Enable LAN services control enables or disables LAN services such as VXI-11, HiSLIP, sockets, and
telnet. However, the LAN interface, home page and configuration page remain enabled for the Web
interface.
Keysight AC6800B Series Operating and Service Guide
41