*PSC 0|1
*PSC?
Enables (1) or disables (0) the clearing of certain enable registers at power on. These include:
Questionable Data Register (STATus:QUEStionable:ENABle)
Standard Operation Register (STATus:OPERation:ENABle)
Status Byte Condition Register (*SRE)
Standard Event Enable Register (*ESE)
The *PSC command does not affect the clearing of the condition or event registers,
just the enable registers. For more information on the SCPI Status System, see Status
Subsystem Introduction.
Parameter
Typical Return
0|1, default 1
0 or 1
Disable power-on clearing of affected registers: *PSC 0
*RCL <0 to 10>
Recalls a saved instrument state.
Parameter
Typical Return
0 to 10
(none)
Recall state from location 1: *RCL 1
This command restores the instrument to a state that was previously stored with the *SAV command.
All instrument settings are recalled except: (1) trigger systems are set to the Idle state, (2) calibration is
disabled, (3) nonvolatile settings are not affected.
This command also aborts TRANsient and ACQuire operations.
Location 0 is recalled at power-on when the Output Power-On state is RCL0.
Stored instrument states are not affected by *RST.
*RST
Resets the instrument to default values. These settings are described in
Parameter Typical Return
(none)
(none)
Reset the instrument: *RST
Keysight AC6800B Series Operating and Service Guide
3 SCPI Programming Reference
Default
Settings.
141