Operating environment
Temperature [°C, (°F)]
Relative humidity [%]
Atmospheric pressure [kPa]
Altitude [m]
1.
Please be aware that the bending angle can be affected if the insertion cord is not kept straight.
2.
Storage under higher temperatures may impact shelf life.
6. Trouble shooting
If problems occur with the system, please use this trouble shooting guide to identify the cause
and correct the error.
Problem
No live image on
the screen but user
interface is present
on the display or
the image is frozen.
Low picture
quality.
10
Possible cause
The endoscope is not connected to
compatible displaying unit.
The displaying unit and endoscope
have communication problems.
The endoscope is damaged.
A recorded image is shown on the
displaying unit screen.
Blood, saliva etc. on the lens (distal tip).
aScope 4 RhinoLaryngo Slim
10 – 40 (50 – 104)
30 – 85
80 – 106
≤ 2000
Recommended action
Connect an endoscope
to the blue port on the
displaying unit.
Restart the
displaying unit.
Replace the endoscope
with a new one.
Return to live image on
the displaying unit.
If the object cannot be
seen clearly, clean the
distal tip.