Télécharger Imprimer la page

NewTom GIANO Prescriptions De Sécurité Et Instructions Pour L'opérateur page 67

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 139
13. INSPECTION AND MAINTENANCE
13.1. USER INSPECTION
These instructions describe the maintenance procedures for the extraoral X-ray system.
These instructions apply to all the revisions of said equipment, as well as all the accessories that may have been provided, therefore the description of some parts may not correspond to your equipment.
Inspection and preventive maintenance must be performed at scheduled intervals to protect the health and safety of patients, users and other persons in accordance with national regulations regarding the use, maintenance of dental x-ray devices
that are in force in the country where the device is installed.
In order to ensure the operational safety and functional reliability of your product, the system owner should check the equipment at regular intervals (at least once a year) or commission an authorized technician to do so.
If one or more checks to be performed are not satisfactory, please contact your dealer tech support.
Step
Check that all labels located
-
on the detector/detectors,
-
on the 3D Panel (if present)
1
-
at the base of the column,
-
on the X-ray tube housing,
are intact, well attached and legible.
Check there are no external damages to the equipment which could reduce protection
2
against radiation.
3
Check the remote emission button cable is not broken or scratched.
Check the power switch verifying that the switch is working properly and the touch panel
4
display illuminates when the switch is in the on position.
5
Verify that exposure is immediately interrupted when X-rays button is released
6
Check the touch console functionality: functions must respond to interrogation.
7
Check proper functioning of X-ray exposure LED and exposure buzzer
8
Check proper functioning of X-ray cenetring laser traces
Check if detector fits easily in the guides Pan and CEPH sides – only for unit with
9
Cephalometric arm and one removable detector.
Perform a motors movement test running the Dummy panoramic procedure, making sure
10
the movement on the 3 axis X, Y,R (rotation and translation simultaneously) is smoothly
and without noise. See figure 1, next page.
In case a Cephalometric arm is installed, perform a motors movement test of the
cephalometric detector using the Dummy Ceph exam procedure, making sure the
11
movement on the axe H (translation) is smoothly and without noise. See figure 1, next
page.
Perform column movements, up and down on the Z axis and check proper functioning.
12
See figure 2, next page.
Check the emergency button functionality. Emergency button is used to stop X-ray
13
device operation, it's located under the patient support arm in proximity of the telescopic
column Emergency button).
Check the X-Ray Generator functionality performing a complete trial exam exposure.
Select any panoramic exam and reset the unit. Hold down the emission button
14
throughout the entire exam procedure. Absence of error messages assures a proper
generator functionality.
If Quality Phantoms are not available at installation site, then contact your dealer tech support to have the Quality Assurance procedure performed at your premises
Perform an exam on the 3D Quality Phantom and evaluate the quality of the outcome
15
volumetric study.
The undersigned confirms that the equipment was checked for the above criteria and that, in case of any malfunction, an authorized technician of the local dealer was informed.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!
EN
OPERATOR'S MANUAL
Description
Section Identification label position
Section Description of operations
Section X-ray emission remote control
Section Starting the system
Section X-ray emission remote control
Section Touch screen display
Section X-ray emission remote control
Section Laser traces
Section Sensor positioning
To execute a dummy run, exam simulation without emitting radiation, select any
panoramic exam and reset the unit. Once the reset position is reached, hold down
button "Reset" throughout the entire exam procedure, otherwise movement will stop.
To execute a dummy run, exam simulation without emitting radiation, select any
cephalometric exam and reset the unit. Once the reset position is reached, hold
down button "Reset" throughout the entire exam procedure, otherwise movement will
stop.
Section Control Panel (console onboard the machine)
User Manual, section Emergency button
Section Performing a 2-D X-Ray examination
Acquisition Operations with Annex
Reference in User Manual
Operator Name
Signature
67
Answer questions with yes (√) or no (−)
Inspection DATE
__/__/20__
__/__/20__
__/__/20__
__/__/20__
__/__/20__

Publicité

loading