Generator reference axis:
10.3. RADIOLOGICAL CHARACTERISTICS IN CB3D MODE
Operating voltage
Generator voltage at maximum current
Load factors for maximum output power
Maximum rated output power
(value averaged with N>4s)
Exposure time (CB3D)
Pulsed mode operating cycle for CB3D exposure
54
90kV with all projections
90kV - 10mA
90kV - 10mA
810VA (90kV 10mA - 20/25% duty cycle)
18s-36s pulsed mode with actual ray total emission time
of 3.6s- 9s
20% - 90%
OPERATOR'S MANUAL
EN