11. ERROR MESSAGES
Code
USER COMMUNICATIONS
WARNING
0.1
0.2
0.3
0.4
0.5
0.6
0.8
0.9
0.10
0.11
0.13
0.14
0.16
EN
Message
The shielded door connected to the
X-ray device is open
X-ray
Button
released
exposure
X-ray Button not released at the end
of examination
Emergency Button pressed
POWER Circuit Board unable to
communicate with MAIN Board
KEYBOARD Circuit Board unable to
communicate with MAIN Board
CONVERTER Circuit Board unable
to communicate with MAIN Board
Circuit Board of CEPH-Detector Lift
(CSENS) unable to communicate
with MAIN Board
Circuit
Board
of
Headrest
(MSENS) unable to communicate
with MAIN Board
Circuit Board of Detectors Carousel
(GSENS) unable to communicate
with MAIN Board
Circuit
Board
of
INVERTER unable to communicate
with MAIN Board
2D Detector failed to turn ON
X-Ray Start-Command aborted
Relocate
2D
Detector
position
OPERATOR'S MANUAL
Solution
Close the door or any other device to
make X-ray emission safe for the
operator.
during
Hold the X-ray emission button until
the end of the procedure.
1) X-ray button was held for too
long after the end of exposure
(>15s).
2) A reset command was sent while
the xray
pressed.
3) The device exited the stand-by
status while the xray button was
being pressed.
Release the emergency stop button.
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
Lift
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
COLUMN-
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
Check
the
contacts. Unplug the sensor and plug
it back.
1) Check
diagnostic LEDs (>15s).
2) Verify CAN connector (KCB1) on
converter board.
to
PAN
Relocate 2D Detector to PAN.
button
was
being
removable
sensor's
Converter
board
63