Message
Relocate 2D Detector to CEPH
position
Check correct position of Craniostat
2D Detector is not recognized (run
"Sensor registration")
2D Detector not calibrated
Serial Number Mismatch: not the
same
2D
Detector
calibration
Detected Open Door (interlock is
active)
Image
not
downloaded
equipment, transfer to PC or USB
before proceeding
Please wait for confirmation before
press
CEPH enabled
CMOS Time and Date Not Set.
CMOS battery low?
Device not configured
Device not calibrated
Please, let the patient release the
bite, then press reset to move the
chinrest to the new position
Buttom pressed on startup
USB Key not found
USB error on writing
Unable to move detector's carousel
automatically.
Please,
manually
OPERATOR'S MANUAL
Solution
Relocate 2D Detector to CEPH
position
Check correct position of Craniostat
Please serialize the sensor (from
unit calibration software on PC)
Calibrate the sensor (from unit
calibration software on PC)
Please serialize the sensor (from unit
used
for
calibration software on PC)
Check interlock switch (door not
closed?)
from
New
exposure
previous has not been downloaded
yet
The xray button was pressed before
starting the cbct exam procedure.
For CBCT exams, launch "Patient
scan" from within NNT and wait for
an explicit request before pressing
the xray button.
Only at first startup after CEPH arm
installation, call tech support if
persistent.
Set time and date from PC/Check if
clock battery on mainboard is dead
Set configuration from PC
Perform unit calibration
Chinrest is going to perform a long
movement, so the patient must
release the bite and let the chinrest
free to move
Detected buttons pressed during
startup. Verify that all the keyboard
buttons are working properly.
USB key not compatible / Corrupted
/ Full
move
it
requested