Symbol on the Keyboard
Tabella 3 – Diagnostic System and Symbols on Keyboard
NOTE: For any error message not listed in the
table, contact a Mectron Service Centre.
Possible Cause
Peristaltic pump malfunction.
Tubing not correctly
positioned in the peristaltic
pump.
The device has been turned
off and on again without
waiting 5 seconds.
Abnormalities in the power
distribution network,
excessive electrostatic
discharges or internal
anomalies.
The system was started with
the pedal pressed.
TROUBLESHOOTING
Solution
Check that there are no
impediments to pump
rotation.
Correctly reposition the
silicon tubing inside the
pump (see
Chapter 4.2 on
page 12
).
Turn device off and wait 5
seconds before turning it on
again.
Turn the device off and wait
5 seconds before turning it
on again.
If the message persists,
contact an Authorized
Mectron Service Center.
Check that during system
start-up the pedal is not
pressed. If the problem
persists, disconnect the
pedal and contact an
Authorized Mectron Service
Center.