Place the insert into the wrench as shown;
Hold the central body of the handpiece
firmly.
CAUTION: Do not grasp the end
and/or cord of the handpiece; only
hold the central body. Do not rotate
the handpiece; simply hold it firmly and
rotate the wrench, only.
Turn the wrench in a clockwise direction
until the clutch engages (the outer part
of the wrench slips, compared to the
handpiece,
producing
clicking sounds "CLICK"). The insert is
now properly tightened;
On the keyboard, select the type of
function, the irrigation level and the light
option, if required.
CAUTION: For the correct
operating parameters to be adopted
for a specific insert, refer to the
Chart annexed to this manual titled
"Appropriate settings for the inserts
on the PIEZOSURGERY® white" or the
illustrative leaflet of the purchased
Mectron insert.
After each use of PIEZOSURGERY® white run a complete cleaning cycle of the irrigation circuit
(PUMP/FLUSH function) using an enzymatic detergent solution, to be followed by a rinse with
distilled water (see the Cleaning and Sterilization Manual).
Switch the device off by means of the O/I switch.
Disconnect all accessories before proceeding with their cleaning and disinfection.
5
6
mechanical
a
7
USE
25