Peristaltic pump tubing
Altitude
Weight and size
Table 2 – Technical Specifications
8.1
Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING: Contraindications.
Interference with other equipment.
Although PIEZOSURGERY® white complies
with standard IEC 60601-1-2, it may interfere
with other equipment located nearby.
PIEZOSURGERY® white must not be used
near, or stacked on other equipment. Install
PIEZOSURGERY® white at a safe distance
from life-support equipment. If necessary,
however, prior their use, please check
and monitor that the device and all other
equipment function correctly in that layout.
WARNING: Portable and mobile radio
communication appliances may affect proper
device operation.
8.1.1
Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
PIEZOSURGERY® white is intended for use in the electromagnetic environment specified below.
The customer or the user of PIEZOSURGERY® white should ensure that it is used in such an
environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
TECHNICAL SPECIFICATIONS
It is advisable do not exceed 8 sterilization
cycles
Less than or equal to 2000 meters
3,2Kg
300 x 250 x 95 mm (L x l x H)
a) W = width ; L = length ; H = height
WARNING: Contraindications.
Interference from other equipment. An
electrical scalpel or other electro-surgical
units near the PIEZOSURGERY® white device
may interfere with its correct operation.
WARNING: The device requires specific
EMC precautions and must be installed
and commissioned according to the EMC
information provided in this chapter.
WARNING: The use of cables and
accessories not supplied by Mectron might
negatively affect the EMC performances.
Compliance
Electromagnetic Environment Guidance
PIEZOSURGERY® white only uses RF
energy for internal function. Therefore,
Group 1
its FR emissions are very low and are not
likely to cause any interference with nearby
electronic equipment.
Class B
PIEZOSURGERY® white is suitable for use
in all establishments, including domestic
Class A
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
Compliant
for domestic purposes.
a)
29