FILLING
AND
CLEANING
IRRIGATION CIRCUIT (Ref. T front
endpaper)
The device is equipped with the "PUMP/
FLUSH" key which, depending on the
mode of use, allows to perform the PUMP
function or the FLUSH function.
The PUMP function can be used at the
beginning of the treatment, to flush the
entire irrigation line up to the insert, so
that the surgery can be started with the
necessary irrigation (see Chapter 5.4 on
page 24).
The FLUSH function allows to perform a
cleaning cycle of the irrigation circuit.
This function must be performed when
you have finished using the device
and before cleaning, disinfecting, and
sterilizing all the parts (see Cleaning and
Sterilization Manual).
THE
function
irrigation
implant
6
5
cortical
4
pump/flush
cancellous
3
special
2
perio
1
endo
0
USE
17