8.1.3
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
8.1.3.1
Power Connection A.C. Input
PIEZOSURGERY® white is designed to operate in the electromagnetic environment specified
below.
The customer or the user of PIEZOSURGERY® white should ensure that it is used in such an
environment.
Basic EMC
Phenomenon
standard or test
Electrical fast
transients /
IEC 61000-4-4
bursts
a) l) o)
Surges
a) b) j) o)
IEC 61000-4-5
Line-to-line
Surges
a) b) j) k) o)
Line-to-
IEC 61000-4-5
ground
Conducted
disturbances
IEC 61000-4-6
induced by RF
fields
c) d) o)
Voltage dips
IEC 61000-4-11
f) p) r)
Voltage
interruptions
IEC 61000-4-11
f) i) o) r)
a) The test may be performed at any one power input
voltage within the PIEZOSURGERY® white RATED
voltage range. If the PIEZOSURGERY® white is tested
at one power input voltage, it is not necessary to re-
test at additional voltages.
Immunity test levels
method
±2 kV on contact
100 KHz repetition
frequency
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2kV
3 V
m)
0.15 MHz - 80 MHz
6 V
in the ISM bands
m)
between 0.15 MHz and
80 MHz
80 % AM at 1 KHz
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycle
Single phase: at 0°
0% UT; 250/300 cycle
TECHNICAL SPECIFICATIONS
Environment Guidance
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF
communication devices
should not be used near any
part of the product, including
cables, except when they
respect the recommended
n)
distances, calculated from the
e)
equation applicable at the
frequency of the transmitter.
g)
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
h)
The quality of the network
voltage should be that of a
h)
typical commercial or hospital
environment.
b) All PIEZOSURGERY® white cables are attached during
the test.
c) Calibration for current injection clamps shall be
performed in a 150 Ω system.
Electromagnetic
31