8.1.4
Specifications of the tests for the Immunity of the Accessible
Parts of the Casing to the Wireless RF Communications Device
PIEZOSURGERY® white is designed to operate in an electromagnetic environment in which
radiated RF disturbances are under control. The customer or the user of PIEZOSURGERY® white
can help prevent electromagnetic interference by ensuring a minimum distance between the
mobile and portable RF (transmitters) communication devices and PIEZOSURGERY® white, as
recommended, in relation to the maximum output power of radiocommunications equipment.
Test
Band
a)
Freq.
(MHz)
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1720
1845
1700 - 1990
1970
2450
2400 - 2750
5420
5500
5100 - 5800
5785
a) For some services, only the uplink frequencies are
included.
b) The carrier shall be modulated using a 50 % duty cycle
square wave signal.
Service
Modulation
a)
Pulse
TETRA 400
modulation
18 Hz
FM
GMRS 460
± 5 kHz
FRS 460
deviation
1 kHz sine
Pulse
LTE band 13,
modulation
17
217 Hz
GSM 800/900
TETRA 800
Pulse
iDEN 820
modulation
CDMA 850
18 Hz
Band LTE 5
GSM 1800
CDMA 1900
GSM 1900
Pulse
DECT
modulation
LTE Band 1, 3,
217 Hz
4, 25
UMTS
Bluetooth
WLAN
Pulse
802.11 b/g/n
modulation
RFID 2450
217 Hz
Band LTE 7
Pulse
WLAN
modulation
802.11 a/n
217 Hz
TECHNICAL SPECIFICATIONS
Max
Distance
power
b)
(W)
1.8
b)
c)
2
0.2
b)
2
b)
2
b)
2
b)
0.2
b)
c) As an alternative to FM modulation, 50 % pulse
modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be
worst case.
Immunity
test level
(m)
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9
35