INSTALLATION
The device must be installed in a suitable and
convenient place for its use.
WARNING: The place where the device is
installed must meet the requirements found
in the
Chapter 4.1 on page 11
PIEZOSURGERY® white can be purchased
4.1
Safety Requirements in the Installation Phase
WARNING: Contraindications.
Interference with other equipment.
Although PIEZOSURGERY® white complies
with standard IEC 60601-1-2, it may interfere
with other equipment located nearby.
PIEZOSURGERY® white must not be used
near, or stacked on other equipment. Install
PIEZOSURGERY® white at a safe distance
from life-support equipment. If necessary,
however, prior their use, please check
and monitor that the device and all other
equipment function correctly in that layout.
WARNING: Contraindications.
Interference from other equipment. An
electrical scalpel or other electro-surgical
units near the PIEZOSURGERY® white device
may interfere with its correct operation.
CAUTION: The electrical system of the
premises in which the device is installed
and used must be compliant to the norms
in force and to the relative electrical safety
precautions.
CAUTION: To avoid any risk of electric
shock this device must be grounded.
WARNING: Do not install the device in
places where there is a risk of explosion.
The device cannot operate in environments
where anesthetic or flammable mixtures are
present.
CAUTION: Position the device in way
so that the power plug is always, easily
reachable, since this plug is considered as a
disconnecting means.
ready for use or may need to be enabled by
typing an activation key.
In case Your device requires an activation key,
the procedures to be followed may vary from
country to country.
.
Always refer to your dealer for details.
peristaltic pump is open, the foot pedal of
the PIEZOSURGERY® white must not be
operated. The parts in movement may cause
harm to the operator.
protected against collisions or against
accidental sprays of water or liquids.
above or near heat sources. Foresee
adequate air circulation around the device
when installing it. Leave adequate space,
especially near the fan placed on the back
part of the device.
CAUTION: Do not expose the device to
direct sunlight or to sources of UV light.
CAUTION: The device can be transported,
but it must be handled with care when it is
displaced. Position the foot pedal on ground,
so that it can only be activated intentionally
by the operator.
CAUTION: Before connecting the
handpiece cord to the device, make sure that
the electrical contacts are perfectly dry. If
need be, dry them with compressed air.
PIEZOSURGERY® irrigation set, check the
integrity of the sterile packaging and inspect
the product to exclude the presence of any
damage. Do not use the irrigation set if the
packaging is open or damaged. The irrigation
set loses sterility if the packaging is broken
or damaged. If the packaging is damaged,
proceed with the disposal of the set. Do not
re-sterilize and reuse the set.
INSTALLATION
WARNING: When the lid of the
WARNING: Install the device in a place
WARNING: Do not install the device
WARNING: Before using the Mectron