Mectron PIEZOSURGERY WHITE Mode D'emploi Et D'entretien page 49

Masquer les pouces Voir aussi pour PIEZOSURGERY WHITE:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Problem
During operation, no liquid
flows out from the insert.
The
device
is
working
properly but the pump works
with difficulty.
The
pump
is
working
properly, but when it stops,
liquid flows out from the
handpiece.
Poor performance
Tabella 4 – Troubleshooting Quick Guide
Possible Cause
The insert is of the type
which does not provide the
passage of liquids.
The insert is clogged.
The handpiece is clogged.
The irrigation level on the
screen is set to "0".
The liquid irrigation bag is
empty
The air inlet on the irrigation
set has not been opened.
The silicon tubing are
incorrectly installed.
Excessive pressure of the
peristaltic pump impeller on
the tube.
The peristaltic pump cover is
not closed properly.
The insert is not correctly
tightened on the handpiece
Broken, worn-out or
deformed insert.
TROUBLESHOOTING
Solution
Use an insert of the type with
the passage of liquids.
Unscrew the insert from
the handpiece and unclog
the insert water passage
by blowing compressed air
through it. If the problem
persists, replace the insert
with a new one.
Contact an Authorised
Mectron Service Centre.
Adjust the irrigation level
Replace the empty bag with
a new one full.
Open the air inlet on the
irrigation set.
Check the irrigation kit
connections.
Check that the tube in the
peristaltic pump has been
correctly inserted.
(Refer to
Chapter 4.2 on
page 12
)
Check that the peristaltic
pump cover is closed
properly.
(Refer to
Chapter 4.2 on
page 12
)
Unscrew the insert and screw
it back correctly by using the
Mectron torque wrench.
(Refer to
Chapter 5.4 on
page 24
)
Replace the insert with a new
one.
39