Delivery
Device Tip
Midline Marker
figure 1: solyx™ sIs system and solyx blue sIs system

WaRnInG

Ensure that the bladder is empty prior to initiating the use of this
product. Ensure that the bladder, urethra and other important
landmarks are properly identified.

steps for Use

Incision and Dissection

1. Prepare and drape the patient using standard surgical
practice. Ensure the bladder is empty.
2. Make a 1.0 cm to 1.5 cm vertical midline incision on the
anterior vaginal wall at the level of the mid-urethra. Dissect
bilaterally to the interior portion of the inferior pubic ramus
at the 45˚ angle off the midline creating a pathway for
delivery device placement.
3. Dissect bilaterally to the interior portion of the inferior pubic
ramus at a 45˚ angle off the midline creating a pathway for
delivery device placement. (See Figure 2).
figure 2: Dissection Pathway

sling Placement

4. Place the mesh assembly onto the delivery device by placing
the delivery device tip into the mesh carrier (See Figure 3).
Note: The mesh carrier should be pushed onto the delivery
device tip until it is flush with the end of the delivery device
shoulder (See Figure 3). While placing the mesh carrier onto
the delivery device, make sure that the mesh is oriented so
that it lies on the outside curve of the delivery device (See
Figure 4). The mesh is now ready for placement.
Black (K) ∆E ≤5.0
Delivery device
Deployment
Mechanism
Mesh Assembly
Mesh Carrier
Polypropylene Mesh
6
Handle
Dissection
Pathway