Type 8025 / SE35, battery powered
Adjustment,Commissioning
8.7.3
Entering or determining the K factor of the fitting used
The K factor of the fitting used is in the Operating Instructions of the fitting.
Find the Operating Instructions of the Bürkert fittings on the internet at www.burkert.com.
The device determines the flow rate in the pipe using the K factor of the fitting.
The K factor of the fitting used can be entered (see Fig. 16) or calculated through a Teach-in procedure (see chap. 8.7.4).
K-FACtOR
FiLtER
Fig. 16 : Entering the K factor of the fitting used
K=10.00
The display shows the last K factor of
the fitting, may it have been entered or
determined by a teach-in procedure.
→
Edit the parameter.
REtURN
K=2.850
→
Enter the K factor (value between 0.010 and
999.9) of the fitting used.
→
Confirm the displayed value.
VALiD N/Y
Is only displayed if the K factor has been changed.
→
Confirm the entered K factor or not.
English