Type 8025 / SE35, battery powered
Adjustment,Commissioning
8
ADjuSTmEnT, commiSSioning
8.1
Safety instructions
Warning
risk of injury due to non-conforming operating.
Non-conforming operating could lead to injuries and damage the
device and its surroundings.
▶ The operators in charge of operating must have read and under-
stood the contents of the Operating Instructions.
▶ In particular, observe the safety recommendations and intended use.
▶ The device/installation must only be operated by suitably trained
staff.
Warning
Danger due to non-conforming commissioning.
Non-conforming commissioning could lead to injuries and damage
the device and its surroundings.
▶ Before commissioning, make sure that the staff in charge
have read and fully understood the contents of the Operating
Instructions.
▶ Before commissioning the device, adjust the K factor. See chap.
8.7.3
▶ In particular, observe the safety recommendations and intended
use.
▶ The device / the installation must only be commissioned by suit-
ably trained staff.
8.2
preparing the device for
adjustment
Switch for the battery
power supply
ON <> OFF
L+ NC L- FE P+ P-
LOCK
Fig. 9 :
Switches on the electronic board
→
Loosen the screw and flip the lid.
→
Loosen the 4 screws and take off the cover of the housing.
→
Put the cover down so that the wires connected to the terminal
block and the wire which connects the flow sensor to the elec-
tronic board are not subject to mechanical stress.
→
Make sure that the ENTER key is unlocked. See Fig. 9.
→
Put the power supply switch to ON. See Fig. 9.
English
Switch to lock/unlock the
ENTER key
LOCK
LOCK
ENTER key
ENTER key
locked
unlocked
15