Adjustment and functions
Adjustmentandfunctions
9.5.2.
entering the K factor or having it determined by teach-in
During the whole teach-in procedure, the outputs are frozen to the status they had at the start of the teach-in.
NO
KFAC
ENTER
ENTER
TEAC
0......9
YES
ENTER
TEAC
To function OUT
* to move the decimal point, simultaneously press keys
49.03
ENTER
KFAC
*
0......9
200.0
ENTER
ENTER
TEAC
VOL
*
0......9
and
0......9
• Entering the K factor of the fitting used:
→
select "NO TEAC",
→
press key "ENTER",
→
In the KFAC parameter, enter the K factor of the
fitting used, in pulse/litre (value between 0,001 and
49.03
ENTER
9999). Find the K factor within the operating instruc-
KFAC
tions of the fitting.
• Determining the K factor by means of a teach-in
procedure:
→
install the 8032 / 8072 in series with a valve, for
instance, then
→
fill a tank with x litres capacity (200 litres for
example).
→
select "YES TEAC",
→
press key "ENTER",
→
open the valve: message "TEAC" blinks.
→
when the tank if full, press again key "ENTER".
→
enter the volume of liquid that has passed through
the circuit (in the unit displayed alternately with the
message "VOL"): the device calculates the K factor
and displays it.
English
33