Type 8202/8222
Fig. 8:
Mounting the display module
Assembly
Check that there is a seal
on the cover and that it is
not damaged. Replace it
if necessary.
G r e a s e t h e s e a l i f
necessary, using a com-
ponent compatible with
the seal material.
Fully tighten by hand to
guarantee tightness.
7.3.
install the ph or redox probe in
the sensor holder of the 8202
(without fluid)
probe head
H
compression washer
seal
A
English
Following instructions
are valid for a Bürkert
probe.
If you use a probe
from another supplier,
respect the related
instructions.
Remove the protective plugs
Check that dimension H on
the probe is between 34
and 46mm. If necessary,
adjust the height of the com-
pression washer.
At first use, apply water or
soapy water on the "A" seal.
Insert the seal into the groove
on the holder.
Insert the probe with its seal
into the holder from above.
17