Type 8025 / SE35, battery powered
Adjustment,Commissioning
8.7.5
configuring the filter
This parameter makes it possible to dampen the fluctuations of the displayed flow rate values.
The filtering has no effect on the totalizers.
Ten filters are available.
When the "fast" filter is active and the flow rate varies for ±30 % (for example when charging the pipe or stopping the flow), the filter
is disabled: the new flow rate is immediately taken into account by the device.
If a filter with a high value is chosen, the device does not detect the sudden flow rate changes thus the flow rate in the pipe may
strongly defer from the displayed flow rate.
A sudden stop of the flow is immediately detected whatever filter is active.
FiLtER
Fig. 19 : Diagram of the "FILTER" parameter of the Parameters menu
FiLtER 0
0......9
FiLtER 9
→
Choose the filter.
→
Confirm
tOtAL
SLOW
0......9
FASt
→
Choose the slow or fast
filter.
→
Confirm