The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an environment.
IEC 60601
IMMUNIT Y TEST
TEST LEVEL
Power frequency
(50 / 60 Hz)
3 A/m
magnetic field. IEC
61000-4-8
Conducted RF IEC
3 Vrms
61000-4-6
3 V/m
Radiated RF IEC
61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot
a
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the ME Equipment is used exceeds the applicable RF compliance level above, the ME Equipment should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the ME Equipment.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
b

Recommended Separation Distances

RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE AND MOBILE RF COMMUNICATION EQUIPMENT AND THE ME EQUIPMENT
The ME Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the ME Equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the ME Equipment as recommended below, according to the maximum output power of the
RATED MAXIMUM OUTPUT POWER OF TRANSMITTER (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
COMPLIANCE LEVEL
Power frequency magnetic fields should be at levels characteristic of typical location in a typical hospital
3 A/m
environment.
Portable and mobile RF communications equipment should be used no closer to any part of the ME Equipment,
including cables, than the recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
3 Vrms
20 V/m
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
the compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with the following symbol:
communication equipment.
150 K Hz to 80 MHz
ELEC TROMAGNETIC
ENVIRONMENT - GUIDANCE
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
.
b
SEPARATION DISTANCE ACCORDING TO FREQUENC Y OF TRANSMITTER (M)
80 MHz to 800 MHz
0.12
0.37
1.17
3.7
11.7
13
800 MHz to 2.5GHz
0.018
0.057
0.18
0.57
1.8
a
, should be less than
0.035
0.11
0.35
1.1
3.5
39296/9527H-0320