Warranty Exclusions; Recommended Separation Distances - Masimo Radius PPG Mode D'emploi

Supplément. avant d'utiliser ce produit, l'utilisateur diot lire et comprendre le mode d'emploi du capteur et ce mode d'emploi
Masquer les pouces Voir aussi pour Radius PPG:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5
FCC ID are as follows: Chip: VKF-AIRTB01
IC IDs are as follows: Chip: 7362A-AIRTB01
Radius PPG Receiver:
EN 301 489-17 V3.1.1
EN 300 328 V2.2.1
EN 301 489-3 V2.1.1
EN 300 330 V2.1.0
WARNING: Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the
equipment.
WARNING: The frequency bands of this device (2.4 GHz) are only for indoor use, in accordance with international telecommunication requirements.
CAUTION: Keep the Radius PPG away from electrical equipment that emits radio frequencies to minimize radio interference. Radio interference may
result in no or inaccurate readings.
Note: This device complies with part 15 of FCC Rules and Industry Canada's license-exempt RSSs' . Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause
undesired operation.
Note: When using Radius PPG consideration should be taken to local government frequency allocations and technical parameters to minimize the
possibility of interference to/from other wireless devices.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio
or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one
or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Consult the dealer or an experienced radio/TV technician for help.
RF Radiation Exposure Statement: This equipment has been exempted from FCC RF radiation exposure testing set forth for an uncontrolled
environment.
Note: Users are advised that high-power radars are allocated as primary users (i.e. priority users) of the bands 5250–5350 MHz and 5650–5850 MHz and
that these radars could cause interference and/or damage to LE-LAN devices
WARRANTY
Masimo warrants to the initial buyer only that these products, when used in accordance with the directions provided with the Products by Masimo, will
be free of defects in materials and workmanship for a period of six (6) months. Single use products are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYER. MASIMO EXPRESSLY
DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR PARTICULAR PURPOSE. MASIMO'S SOLE OBLIGATION AND BUYER'S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE,
AT MASIMO'S OPTION, TO REPAIR OR REPLACE THE PRODUCT.

WARRANTY EXCLUSIONS

This warranty does not extend to any product that has been used in violation of the operating instructions supplied with the product, or has been subject
to misuse, neglect, accident or externally created damage. This warranty does not extend to any product that has been connected to any unintended
device or system, has been modified, or has been disassembled or reassembled. This warranty does not extend to products that have been reprocessed,
reconditioned or recycled.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR CONSEQUENTIAL
DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE POSSIBILITY THEREOF. IN NO EVENT SHALL MASIMO'S
LIABILITY ARISING FROM ANY PRODUCTS SOLD TO BUYER (UNDER A CONTRACT, WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT PAID
BY BUYER FOR THE LOT OF PRODUCT(S) INVOLVED IN SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES ASSOCIATED
A PRODUCT THAT HAS BEEN REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL NOT BE DEEMED TO
PRECLUDE ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDED BY CONTRACT.
NO IMPLIED LICENSE
PURCHASE OR POSSESSION OF THIS PRODUCT DOES NOT CARRY ANY EXPRESS OR IMPLIED LICENSE TO USE WITH ANY DEVICE THAT IS NOT AN
AUTHORIZED DEVICE OR SEPARATELY AUTHORIZED TO USE MASIMO PRODUCTS.

RECOMMENDED SEPARATION DISTANCES

RECOMMENDED SEPARATION DISTANCE BETWEEN PORTABLE AND MOBILE RF COMMUNICATION EQUIPMENT AND THE ME EQUIPMENT
The ME Equipment is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the ME Equipment can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the ME Equipment as recommended below,
according to the maximum output power of the communication equipment.
RATED MAXIMUM OUTPUT POWER
OF TRANSMITTER (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
EU Radio Equipment Directive (RED 2014/53/EU)
SEPARATION DISTANCE ACCORDING TO FREQUENC Y OF TRANSMITTER (M)
150 kHz to 80 MHz
d = 1.17*√P
0.12
0.37
1.17
3.7
11.7
Radius PPG Chip:
EN 301 489-17 V3.1.1
EN 300 328 V2.2.1
80 MHz to 800 MHz
800 MHz a 2.5 GHz
d = 1.17*√P
d = 2.33*√P
0.12
0.23
0.37
0.74
1.17
2.33
3.7
7.37
11.7
23.3
3
10693B-eIFU-0420

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Safetynet

Table des Matières