SYMPTOM
Battery run time significantly
reduced
Battery not charging after plugged
into AC power source
SPECIFICATIONS
Display Range
MEASUREMENT
SpO
(Functional Oxygen Saturation)
2
PR (Pulse Rate)
Pi (Perfusion Index)
RRp (Respiration Rate from the Pleth)
The emitted wavelengths range from 600 nm to 1000 nm and the peak optical power is less
than 15 mW. Information about the wavelength range can be especially useful to clinicians.
Accuracy (A
)
rms
OX YGEN SATURATION (SpO
No Motion [1] (SpO
from 70% to 100%)
2
Motion [2] (SpO
from 70% to 100%)
2
Low perfusion [3] (SpO
from 70% to
2
100%)
RESPIRATORY RATE (RRP) [5], [6], [7]
Range of 4 to 90 RPM
* A
accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/- A
rms
the reference measurements in a controlled study.
Note: A functional tester cannot be used to assess the accuracy of Rad-G.
Electrical
AC POWER REQUIREMENTS
AC Power requirements
Power consumption
Environmental
RAD-G DEVICE ENVIRONMENTAL CONDITIONS
Operating Temperature
Storage/Transport Temperature
Operating Humidity
Storage/Transport Humidity
Operating Atmospheric Pressure
*Exceeding this temperature can cause charging to stop.
**Compliance with IEC 60601-1 surface temperature requirements evaluated at 40°C.
POTENTIAL CAUSES
• Battery not fully charged.
• Battery damaged.
• Battery capacity effected.
• Battery is completely depleted.
DISPLAY RANGE
0% to 100%
25 BPM to 240 BPM
0.00 to 20.0
4 RPM to 90 RPM
)
2
Adults, Pediatrics, Infants
Adults, Pediatrics, Infants
Adults, Pediatrics, Infants
1 RPM
100-240VAC, 50/60Hz, 0.6A
< 6W
While battery is charging*: 0°C to 40°C (32°F to 104°F)
While battery is NOT charging: 0°C to 50°C**
(32°F to 122°F)
-20°C to 60°C (-4°F to 140°F) [9]
10% to 95%, non-condensing
10% to 95%, non-condensing
540 mbar to 1060 mbar (540 hPa to 1060 hPa)
• Check battery charge level indicator.
• Check battery is fully charged.
• Contact Masimo service. (See Contacting Masimo section).
• Check the Rad-G is ON during recharging if the battery is completely depleted.
• Contact Masimo service. (See Contacting Masimo section).
Resolution
PARAMETER
SpO
(Functional Oxygen Saturation)
2
PR (Pulse Rate)
Pi (Perfusion Index)
RRp (Respiration Rate)
Range
2%
No Motion
3%
Motion [4]
2%
Low perfusion
Battery
BATTERY
Type
Capacity
Charging Time
*Time to reach 80% capacity at 25°C ambient temperature.
Physical Characteristics
PHYSICAL CHARAC TERISTICS
Dimensions
Weight
11
NEXT STEPS
RESOLUTION
1%
1 BPM
0.01
1 RPM
PULSE RATE (PR)
25 BPM to 240 BPM
Adults, Pediatrics, Infants
Adults, Pediatrics, Infants
Adults, Pediatrics, Infants
Lithium ion
24 hours [8]
8 hours*
7.4 cm x 19.8 cm x 2.5 cm
(2.9" x 7.8" x 1.0")
0.27 kg. (0.59 lbs.)
3 BPM
5 BPM
3 BPM
of
rms