4 OPERATION
4.3 Calibration Mode: Press
The following adjustments are set in the calibration mode display:
LANGUAGE
UNIT
CELL CST
T ° COEFF
CURRENT
RELAY
FILTER
END
4.3.1 Language
ENTER
LANGUAGE
UNIT
ENTER
4.3.2 Engineering Units
The conductivity can be displayed in any engineering unit with 0, 1, 2 or 3 decimal points
but always with 4 digits.
Choice TDS (Total Disolved Solids) makes it possible to display the concentration in ppm.
Enter KTDS = TDS (ppm) / Conductivity (µS/cm) = 0,46 (NaCl)
ENTER
Selection between English, German, French or Italian.
Selection of engineering units for conductivity and temperature.
Enter the sensor cell constant
Selection of temperature compensation mode (linear, automatic
or Teach-in function).
Determination of 4-20 mA measuring range.
Parameter definition of relays. This message only appears if the
relay option has been installed.
Damping selection. There are ten different steps available.
Back to the operation mode and save the new parameters.
ENGLISH
DEUTSCH
FRANCAIS
ITALIANO
E-35-
CONDUCTIVITY 8225
simultaneously for 5 seconds
The selected language is confir-
med by the Enter-key and becomes
immediately active.