To Test The Pumping System - Agilent Technologies TPS-compact Mode D'emploi

Masquer les pouces Voir aussi pour TPS-compact:
Table des Matières

Publicité

WARNING!
If dangerous gases were being pumped, flush the pump with air or inert gas
for at least 10 minutes prior to disassembly.

To test the Pumping System

1
2
3
Newly installed tip seals may require several hours of run time to seat properly
NOTE
and enable the pump to meet speed and base pressure specifications.
TPS-compact User Manual / 87-901-010-01 (H)
Run the TPS-compact for about 5 seconds. Verify that the front
fan is running. If you hear loud noises from the Scroll pump this
indicates that the tip seal or main O-ring are possibly out of
place.
Disassemble and repair as necessary.
The pump is now ready to return to service.
Technical Information
Service Options
305/314
17