It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
The Navigator Software (optional) allows the operator to set all the
NOTE
programmable feature.
How to Connect the Open Collector Input of the
Controller
Here below there are the typical connections of the open collector
input of TPS-compact to an external system. Two cases are
considered:
TPS-compact User Manual / 87-901-010-01 (H)
reference quantity: frequency
thereshold: 867 Hz
hysteresis: 2 %
activation type: high level
delay time: 0 second
Technical Information
Interconnections