o If the test is not successful, an audible indication does not sound. See 'Troubleshooting'
section of this manual for appropriate steps.
9. Perform the localization procedure by applying the Handpiece tip to skin or soft tissue.
Ensure that no air gaps are present between the Handpiece tip and tissue.
When the Handpiece detects the Reflector, the Console will emit an audible
feedback that increases in cadence as the Handpiece is placed closer to the
Reflector. The distance reading will decrease as the handpiece is placed closer to
the Reflector.
NOTE: When not in use, place the Console in Standby mode by pressing the Start/Stop button.
The display will show SYSTEM READY. To return to LOCALIZATION MODE, perform Step 7 again.
After Surgery
10. Turn off the Console by pressing the power switch off ( O ).
11. Remove the Handpiece from the front panel
12. Unplug the power cord from the wall outlet.
CLEANING
Warning
• Electric Shock Hazard – Always turn off and unplug the Console before cleaning.
Notice
• Do not clean the Console with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the console.
1. Turn off the Console, and unplug the power cord from the wall outlet.
2. Thoroughly wipe all surfaces of the Console, including handpiece receptacles, and power
cord with a mild cleaning solution or disinfectant and a damp cloth.
Follow universal, generally accepted practices when handling components that have come in
contact with blood or tissue.
Follow the procedures approved by your institution or use a validated infection control
procedure. Do not allow fluids to enter the Console. The Console cannot be sterilized.
TROUBLESHOOTING
General Troubleshooting Guidelines
If the Console malfunctions, check for conditions that may have caused the problem:
• Check the Console for visible signs of physical damage.
• Verify that the power cord is connected to the Console and power outlet.
• Verify that the Handpiece cable is connected and attached properly.
Use the table below to help identify and correct specific conditions.
Situation
Recommendation
Display does not light up
• Verify that AC power cord is connected.
• Verify that the AC power outlet is live.
• Replace power cord.
• Replace power fuse (replace with a T 1AL, 250V type fuse).
SYSTEM READY message
• Verify that the Handpiece is connected and attached properly.
does not appear
• Turn the Console off, then on again.
• Replace Handpiece.
• If symptom persists, the Console may require service. Contact
Merit Medical.
CH ECK CO N N EC T I O N S
• Completely detach Handpiece and re-attach to ensure fully
message appears
connected.
• Verify that the Handpiece connections are connected and
attached properly.
• Turn the Console off, then on again.
• Replace Handpiece.
• If symptom persists, the Console may require service. Contact
Merit Medical.
N o s o un d h eard f ro m
• Verify that the console is in LOCALIZATION MODE.
Console / visual display
• Verify that the Handpiece connections are connected and
numbers do not change
attached properly.
• Turn the Console off, then on again.
• Replace Handpiece.
• Use self-test card to check for system function.
•If symptom persists, the Console may require service. Contact
Merit Medical.
R E P L A C E H A N D P I E C E
• Handpiece has expired and will need to be replaced.
m e ss a g e a p p e a r s (f o r
• Turn power off, replace handpiece then turn power on.
disp os ab l e han dp ie ce
only)
LO C A L I Z AT I O N M O D E
• Press START/STOP button
message does not appear
• Turn the Console off, then on again.
• If symptom persists, the Console may require service. Contact
Merit Medical.
Clicking during nonuse of
• Press START/STOP button, put system in SYSTEM READY
handpiece
• Ensure console is not stacked on top of, or under, other
equipment.
• Move handpiece away from other cables e.g. electrocautery
system cable
• Turn the Console off, then on again.
MAINTENANCE AND SERVICE
The SCOUT Console must be serviced by Merit Medical, Inc. To ensure proper performance,
the Console should be checked for any signs of damage or wear before each use and storage.
This includes the housing, integrity of switches, and power cord. Should abnormalities be
discovered by the user, the user should contact Merit Medical, Inc. Do not use a damaged
Console.
Preventative maintenance is limited to external cleaning of console, fuse replacement, and
confirmation of power up functional diagnostics as described in this manual. The SCOUT
console does not require any specific routine service.
The Console contains no user serviceable parts and should not be opened by the user. Please
contact Merit Medical for service.
Merit Medical Systems, Inc.
6 Journey, Suite 125
Aliso Viejo, CA 92656 USA
www.merit.com
Caution
• If a Console is to be shipped from your institution for repair, then please clean and disinfect the
Console as described in this manual before packing for shipment. Indicate on the outside of the
shipping container that the Console has been cleaned and disinfected.
SYMBOL
Serial Number
Catalog Number
Warning: risk of electric shock. No user serviceable parts inside.
See instructions for use
General warning
Manufacturer of the device
Identifies a type BF applied part
Caution: Consult accompanying documents
Medical Device.
"For European Union (EU) States, this symbol indicates 'Not for general waste.'
Dispose of in accordance with the waste electronic & electrical equipment
(WEEE) directive."
SPECIFICATIONS
Parameter
Operating power
Power consumption
Audio volume
Sound indicator
Visual Indicators
Instrument make / model
Dimensions
Weight
Operating ambient temperature range
Operating relative humidity range
Storage ambient temperature range
Storage relative humidity range
Shipping temperature:
Shipping humidity:
Storage and transit atmospheric pressure
Operating altitude
System Controls
Interconnects
Ingress of Fluids
Flammability
DESIGNATION
Specification
AC Line Power 120/240 VAC (50/60Hz)
26 watts, nominal
38 watts, maximum (at high volume)
70 dB Sound Pressure Level at 1 meter
Frequency proportional to event rate.
LCD display with system messages
Merit Medical Inc. / SCOUT Console
12 inches width
6 inches height
12 inches depth
6 lbs.
10°C to 30°C (50° to 86° F)
30% to 75%, non-condensing
-20o to 60o C (-40o to 104o F)
10 percent to 95 percent relative humidity,
non-condensing
- 4 0 ° C to 6 0 ° C (- 4 0 ° F to 14 0 ° F) for 3
consecutive days.
10% to 95% RH, non-condensing, for 3
consecutive days.
500hPa to 1060 hPa (7.3 psia to 15.4 psia)
Maximum of 2000m
Buttons
Custom cables
Not protected against the ingress of fluid.
IPX0
Equipment not suitable for use in the
presence of flammable anesthetic mixture
with air or with oxygen or with nitrous oxide