3
Features and Functions
Carrier Frequency
The maximum carrier frequency depends on the function selected as shown
below. The default is 1 kHz for all functions.
– The carrier frequency must always be greater than or equal to the frequency
– The sum of the carrier frequency and deviation must be less than or equal to
– Front-Panel Operation: To set the carrier frequency, press the Freq softkey for
– Remote Interface Operation:
96
Function
Sine
Square
deviation. If you attempt to set the deviation to a value greater than the carrier
frequency (with FM enabled), the function generator will automatically adjust
the deviation to the maximum value allowed with the present carrier
frequency.
the maximum frequency for the selected function plus 100 kHz (10.1 MHz for
sine and square waveforms). If you attempt to set the deviation to a value that
is not valid, the function generator will automatically adjust it to the maximum
value allowed with the present carrier frequency.
the selected function. Then use the knob or numeric keypad to enter the
desired frequency.
FREQuency {<frequency>|MINimum|MAXimum}
You can also use the APPLy command to select the function, frequency,
amplitude, and offset with a single command.
Minimum Frequency
1 mHz
1 mHz
Maximum Frequency
10 MHz
10 MHz
Keysight 33210A User's Guide