Télécharger Imprimer la page

Keysight 33210A Guide D'utilisation page 180

Générateur de signaux arbitraires/fonctions 10 mhz

Publicité

4
Remote Interface Reference
Simplified Programming Overview
This section gives an overview of the basic techniques used to program the
function generator over the remote interface. This section is only an overview and
does not give all of the details you will need to write your own application
programs. Refer to the remainder of this chapter and also the application
examples in
reference manual that came with your programming application for further details
on instrument control.
Using the APPLy Command
The APPLy command provides the most straightforward method to program the
function generator over the remote interface. For example, the following
command string sent from your computer will output a 3 Vpp sine wave at 5 kHz
with a -2.5 volt offset.
Using the Low-Level Commands
Although the APPLy command provides the most straightforward method to
program the function generator, the low-level commands give you more flexibility
to change individual parameters. For example, the following command strings
sent from your computer will output a 3 Vpp sine wave at 5 kHz with a -2.5 volt
offset.
180
Chapter 6
for more details. You may also want to refer to the
APPL:SIN 5.0E+3, 3.0, -2.5
FUNC SIN
FREQ 5000
VOLT 3.0
VOLT:OFFS -2.5
Select sine wave function
Set frequency to 5 kHz
Set amplitude to 3 Vpp
Set offset to -2.5 Vdc
Keysight 33210A User's Guide