Calibration Message
The function generator allows you to store one message in calibration memory in
the mainframe. For example, you can store such information as the date when the
last calibration was performed, the date when the next calibration is due, the
function generator's serial number, or the name and phone number of the person
to contact for a new calibration.
– You can record a calibration message only from the remote interface and only
– The calibration message may contain up to 40 characters (additional
– Storing a calibration message will overwrite any message previously stored in
– The calibration message is stored in non-volatile memory and does not change
– Front-Panel Operation: Press
– Remote Interface Operation: To store the calibration message, send the
160
when the function generator is unsecured. You can read the message from
either the front-panel or over the remote interface. You can read the
calibration message whether the function generator is secured or unsecured.
characters are truncated).
memory.
when power has been off or after a remote interface reset.
the "Test / Cal" menu. The calibration message is listed as one of the messages
on the display.
following command, including any desired string within the single quotes
(' ').
CAL:STR 'Cal Due: 01 August 2008'
and then select the Cal Info softkey from
Keysight 33210A User's Guide