Marker Frequency
If desired, you can set the frequency at which the signal on the front-panel
Sync connector goes to a logic low during the sweep. The Sync signal always
goes from low to high at the beginning of the sweep.
– Marker frequency: 1 mHz to 10 MHz (limited to 100 kHz for ramp waveforms).
– When the sweep mode is enabled, the marker frequency must be between the
– The Sync enable setting is overridden by enabling the marker used with the
– Front-Panel Operation: After enabling sweeps, press the Marker softkey. Then
– Remote Interface Operation:
Keysight 33210A User's Guide
The default is 500 Hz.
specified start frequency and stop frequency. If you attempt to set the marker
frequency to a frequency not in this range, the function generator will
automatically set the marker frequency equal to the start frequency or stop
frequency (whichever is closer).
sweep mode (see page 83). Therefore, when the marker is enabled (and the
sweep mode is also enabled), the Sync setting is ignored.
use the knob or numeric keypad to enter the desired marker frequency.
MARKer:FREQuency {<frequency>|MINimum|MAXimum}
MARKer {Off|On}