Error Messages
5
-221
Settings conflict;
burst period increased to fit entire burst
The number of cycles specified by the BURS:NCYC command takes priority over
the burst period (as long as the burst period is not at its maximum value). The
function generator has increased the burst period to accommodate the specified
burst count or waveform frequency.
-221
Settings conflict;
burst count red uced to fit entire burst
Since the burst period is currently at its maximum, the function generator has
reduced the burst count to allow the specified waveform frequency.
-221
Settings conflict;
triggered burst not available for noise
You cannot use the noise function in the triggered burst mode.
-221
Settings conflict;
amplitude units changed to Vpp d ue to high-Z load
The output units (VOLT:UNIT command) cannot be set to dBm if the output
termination is currently set to "high impedance" (OUTP:LOAD command). The
function generator has converted the units to Vpp.
-221
Settings conflict;
trigger output connector d isabled by trigger external
When the external trigger source is selected (TRIG:SOUR EXT command), the
function generator automatically disables the "trigger out" signal. The rear-panel
Trig connector cannot be used for both operations at the same time.