Output Units
Applies to output amplitude only. At power-on, the units for output amplitude are
volts peak-to-peak.
– Output units: Vpp, Vrms, or dBm. The default is Vpp.
– The unit setting is stored in volatile memory. The units are set to "Vpp" when
– The function generator uses the current units selection for both front panel and
– The output units for amplitude cannot be set to dBm if the output termination
– Front-Panel Operation: Use the numeric keypad to enter the desired
– Remote Interface Operation:
Keysight 33210A User's Guide
power has been off or after a remote interface reset (provided the Power On
state is set to "default").
remote interface operations. For example, if you select "VRMS" from the
remote interface, the units are displayed as "VRMS" on the front panel.
is currently set to "high impedance". The units are automatically converted to
Vpp.
magnitude and then press the appropriate softkey to select the units. You can
also convert from one unit to another from the front panel. For example, to
convert 2 Vpp to its equivalent value in Vrms, press
V
softkey. The converted value is 707.1 mVrms for a sine wave.
RMS
VOLTage:UNIT {VPP|VRMS|DBM}
Features and Functions
and then press the
3
77